| Model Number |
LSP201A |
| Medical Device Problem Codes |
Low impedance (2285); Capturing Problem (2891)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226); Cardiac Perforation (2513)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during the implant procedure of a right atrium leadless pacemaker (ralp) that after fixating and transitioning to tether mode that there was low pacing impedance and high capture threshold.Angiography was performed and revealed that the helix of the ralp perforated the right atrial appendage.After some time, the physician elected to unscrew since the threshold was not decreasing, at this time, the patient's blood pressure dropped due to cardiac tamponade.The physician elected to abandon the implant procedure.The physician performed pericardiocentesis and the blood pressure stabilized.After about an hour the patient worsened again, and physician decided to open the chest and perform surgery to repair the wound.The patient was stable following the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
|
| |
|
Search Alerts/Recalls
|