| Model Number |
L311 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker and the right atrial (ra) lead exhibited oversensing of non-physiological and mild myopotential noise.This pacemaker recorded an inappropriate atrial tachy response (atr) episode due to the oversensing.Technical services (ts) discussed performing isometrics to test for noise and impedance fluctuations.Also, ts provided reprogramming options.This pacemaker remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|