BOSTON SCIENTIFIC CORPORATION ACCOLADE? MRI DR; Implantable Pulse Generator, Pacemaker (Non-Crt) Back to Search Results
Model Number L311
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this pacemaker and the right atrial (ra) lead exhibited oversensing of non-physiological and mild myopotential noise.This pacemaker recorded an inappropriate atrial tachy response (atr) episode due to the oversensing.Technical services (ts) discussed performing isometrics to test for noise and impedance fluctuations.Also, ts provided reprogramming options.This pacemaker remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE? MRI DR
Common Device Name
Implantable Pulse Generator, Pacemaker (Non-Crt)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50935898
Report Number2124215-2026-51742
Device Sequence Number18516235
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public(01)00802526559228(17)260221
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P150012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/21/2026
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number771794
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured02/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexMale
-
-