ABBOTT VASCULAR INC. XIENCE Skypoint¿; Coronary Drug-Eluting Stent Back to Search Results
Model Number 1804350-18
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat the circumflex coronary artery.The 3.5x18 mm xience skypoint stent dislodged from the delivery system prior to reaching the target lesion.There was no resistance.The stent went down into the aorta and is undeployed.A snare was attempted but was unsuccessful.A non-abbott stent was used to treat the target lesion and the patient was discharged as planned.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, a definitive cause for the reported issue could not be determined.Stent dislodgement can be influenced by several factors, including, but not limited to, improper or inadequate stent crimping during manufacturing, incorrect sheath sizing, positive pressure applied during device preparation, forced sheath removal, mishandling of the stent during preparation, and interactions with patient anatomy and/or other interventional devices.An attempt to retrieve the dislodged stent using a snare was unsuccessful, and an additional stent was subsequently implanted to treat the target lesion.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.C3: 1 per year was entered; however, the frequency is 1 dose per implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE Skypoint¿
Common Device Name
Coronary Drug-Eluting Stent
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005718570 (P099)
cashel road
clonmel tipperary UNK
EI   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50935930
Report Number2024168-2026-03753
Device Sequence Number13161239
Product Code NIQ
UDI-Device Identifier08717648233401
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1804350-18
Device Catalogue Number1804350-18
Device Lot Number5052641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient Weight214 LB
-
-