The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, a definitive cause for the reported issue could not be determined.Stent dislodgement can be influenced by several factors, including, but not limited to, improper or inadequate stent crimping during manufacturing, incorrect sheath sizing, positive pressure applied during device preparation, forced sheath removal, mishandling of the stent during preparation, and interactions with patient anatomy and/or other interventional devices.An attempt to retrieve the dislodged stent using a snare was unsuccessful, and an additional stent was subsequently implanted to treat the target lesion.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.C3: 1 per year was entered; however, the frequency is 1 dose per implant.
|