• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Sistemas Medicos Alaris S.A de C.V. BD Alaris Extension Set; Intravascular Administration Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Sistemas Medicos Alaris S.A de C.V. BD Alaris Extension Set; Intravascular Administration Set Back to Search Results
Catalog Number 11088483
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd alaris extension set had damaged tubing.The following information was provided by the initial reporter: we are still experiencing tubing that is ¿breaking.¿ our clinical staff in cw has most recently reported the following: 4 extension set: umid 1609 mfg# 11088483 additional information received from the customer: was the fault observed before priming, during priming or after priming the set? most are after priming and standard use, some occur during priming please confirm the exact location of the reported fault within the set? at the leur connection did it fall apart at where the tubing connects? yes what was the impact to the patient? loss of drug; risk of infection can you please provide an exact date of event in format dd-mmm-yyyy? too many dates to provide.Risk reports were created can you please provide the lot number associated with reported issue? i do not have specific lot numbers as this has occurred over a several month period and the packaging has been thrown out as part of setting up the tubing.Can you please confirm all 4 extension set involve same patient or different patients? different patients.
 
Additional Manufacturer Narrative
B.3.The date of event is unknown; bd awareness date used.E.1.Address was not located, and il was used.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Alaris Extension Set
Common Device Name
Intravascular Administration Set
Manufacturer (Section D)
Sistemas Medicos Alaris S.A de C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX  22250
Manufacturer (Section G)
Sistemas Medicos Alaris S.A de C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
+1(847)393-5694
MDR Report Key50935980
Report Number9616066-2026-02468
Device Sequence Number18831132
Product Code FPA
UDI-Device Identifier10885403234668
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K790108
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number11088483
Device Lot NumberUnknown
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-