| Model Number |
M00539230 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Difficult to Remove (1528); Device-Device Incompatibility (2919); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a flexima biliary preloaded stent was to be implanted during a stent placement procedure performed on an unknown date.During the procedure, the guidewire ruptured inside the introducer, resulting in difficulty actuating and properly deploying the flexima biliary stent.The stent was deployed at an incorrect location and was not possible to be removed and additional measures were required to address the event.Another of the same device was used to complete the procedure.The patient the experienced hemodynamic instability, progressing to a drop in hb, and the patient subsequently died.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3:
approximated based on the date the manufacturer became aware of the event.Block h6:
imdrf device code a150207 captures the reportable event of stent difficult to remove.Imdrf patient code e2401 captures the reportable event of hemodynamic instability, progressing to a drop in hb.Imdrf impact code f02 captures the reportable event of death.
|
| |
|