| Catalog Number |
K3186N24E |
| Medical Device Problem Codes |
Fracture (1260); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer states the armrests have stripped screws and the brakes are not holding on both sides.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer states the armrests have stripped screws and the brakes are not holding on both sides.Sample and/or photo provided? not available; pcn history reviewed and no product changes were identified that could have contributed to the alleged product failure; quality control inspections records reviewed and the result was passed inspection.Investigation results: not confirmed.Sample required to determine cause; the division will monitor through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|
Search Alerts/Recalls
|
|
|