ICU MEDICAL DE MEXICO, S. De R.L. De C.V. 120" (305 cm) Appx 15.8 ml, 15 Drop Admin Set w/3 Clave¿ Clear, NanoClave¿ Stopc; Stopcock, I.V. Set Back to Search Results
Catalog Number AC135
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
Additional used device has been received for the investigation, and the event involved a 120" (305 cm) appx 15.8 ml, 15 drop admin set with 3 clave clear, nanoclave stopcock, 2 check valves, spin luer, 2 ext.It was reported that set was primed successfully but would not deliver fluid when connected to the patient¿s catheter.Saline was involved.There was no delay in therapy because another set was promptly primed and used.There was patient involvement and no patient harm was reported.
 
Additional Manufacturer Narrative
One used device was returned for evaluation.As received, no physical damage or other anomalies were observed on the returned samples.The sets and each of the ports were primed according to procedure, and no occlusions or other anomalies were observed along the devices.The reported condition of no flow, inability to prime, or difficulty priming could not be confirmed or replicated based on the physical sample evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
120" (305 cm) Appx 15.8 ml, 15 Drop Admin Set w/3 Clave¿ Clear, NanoClave¿ Stopc
Common Device Name
Stopcock, I.V. Set
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. De R.L. De C.V.
avenida cuarzo no. 250
ensenada, b.cfa. NA 22790
MX  22790
Manufacturer Contact
reed covert
951 calle amanecer
san clemente, CA 92673
+1(224)706-2300
MDR Report Key50936392
Report Number9617594-2026-01366
Device Sequence Number9589365
Product Code FMG
UDI-Device Identifier00887709124417
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K964435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAC135
Device Lot Number14991102
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-