Abbott Medical (CRM-Sylmar) QUARTET; Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) Back to Search Results
Model Number 1458Q/86
Medical Device Problem Codes Failure to Capture (1081); Separation Failure (2547); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that loss of capture was noted on the left ventricle (lv) lead.The device was reprogrammed to deactivate the lv lead.Later, lead migration occurred during a revision procedure.Repositioning using a guidewire was attempted, however, the guidewire was unable to release from the lv lead.The lv lead was explanted and replaced.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Common Device Name
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50936437
Report Number2017865-2026-20633
Device Sequence Number9563811
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public(01)05414734503198(10)A000188386(17)290531
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1458Q/86
Device Lot NumberA000188386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/06/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-