| Model Number |
18320 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Hyperglycemia (1905)
|
| Date of Event |
08/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the legal guardian that the patient's blood glucose level rose to 330 mg/dl while wearing the pod between 5 and 24 hours.When removed from the infusion site (abdomen), the pod's cannula was found to be dislodged and the pod was leaking insulin.The patient also experienced some bleeding.As treatment, a new pod was applied.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed, and the product lot met all acceptance criteria.
|
| |
|
Search Alerts/Recalls
|