• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Insulet Corporation Omnipod 5 Pod; Automated Insulin Delivery System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Insulet Corporation Omnipod 5 Pod; Automated Insulin Delivery System Back to Search Results
Model Number PT-001444
Medical Device Problem Codes Loss of or Failure to Bond (1068); Fluid/Blood Leak (1250); Material Twisted/Bent (2981)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the patient's blood glucose level rose to 400 mg/dl while wearing the pod between 4 and 24 hours.The pod was leaking, causing the adhesive to peel up, and when the pod was removed from the infusion site (leg), the pod's cannula was found bent.As treatment, a new pod was applied.
 
Additional Manufacturer Narrative
The device has been returned; however, the investigation is not yet complete.When the investigation is complete, a supplemental report will be submitted with the investigation results.Lot release records were reviewed and the product lot met all acceptance criteria.Locked down smartphone: lockdown omnipod software app version: 4.0.2 operating system: n5004l-am-q-mv01602-06-01.06 hardware: n5004l cgm sensor type: data not available please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod 5 Pod
Common Device Name
Automated Insulin Delivery System
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50936686
Report Number3014585508-2026-72077
Device Sequence Number9563814
Product Code QFG
UDI-Device Identifier20385083000425
UDI-Public(01)20385083000425(11)260112(17)280112(10)PH1U01122631
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K231826
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPT-001444
Device Catalogue NumberPOD-OMNI-I1-6220
Device Lot NumberPH1U01122631
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/21/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured01/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-