| Model Number |
105-5081-153 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Medtronic received a report that a solitaire ab stent was damaged during the procedure and encountered resistance the patient was undergoing surgery for treatment of a m1 ischemic stroke.The mrs baseline and procedure score was 1.The nihss baseline and post procedure score was 1.The tici score baseline and post procedure was 0.Iv tpa (intravenous tissue plasminogen activator) was not contraindicated.There were 2 passes with the device.It was noted the patient's vessel tortuosity was normal.Stroke onset to reperfusion time was 2 hours.It was reported that the patient had acute cerebral infarction and underwent thrombectomy in the right middle cerebral artery m1 segment.During the procedure, the sab-6-30 was used for thrombectomy.During delivery through the middle segment of the rebar18, the stent became stuck in the microcatheter.After the stent was removed, it was observed that the tail end of the stent had a kink, and the tail end of the stent was severely deformed.After the stent was replaced, the procedure proceeded smoothly.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.Ancillary devices include a rebar 18 microcatheter.Additional information was received reporting that no causes or contributing factors for the event were identified.Catheter resistance occurred in the middle of the device.A continuous saline flush was administered and maintained as indicated in the ifu.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|