| Catalog Number |
03.037.018 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative:
d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the stepped ream for tfna helical blade and screw was used during insertion of a tfna nail.During preparation for the proximal locking portion of the case, the 10 mm cannulated drill bit was used without incident, after which the 6/9 tapered drill bit fractured at the tip in the femoral neck and was described as crumbling like dust after brief contact in the patient.It was also reported that substantial torque was applied to the insertion handle and aiming arm jig to achieve trajectory and reduction, and that the 3.2 guide wire bent during insertion of the lag screw head element.Additional impacted products were the insert handle, 3.2 mm wire guide sleeve, lock insert for aim arm, 3.2 mm trocar, blade/screw guide sleeve, and 125 degree aiming arm.Fragments were generated, retained in the patient, and additional intervention was required to remove the retained fragments.
|
| |
|
Search Alerts/Recalls
|
|