• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Roche Diagnostics BILT3 Bilirubin Total Gen.3; DIAZO COLORIMETRY, BILIRUBIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Roche Diagnostics BILT3 Bilirubin Total Gen.3; DIAZO COLORIMETRY, BILIRUBIN Back to Search Results
Catalog Number 10421436190
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BILT3 Bilirubin Total Gen.3
Common Device Name
DIAZO COLORIMETRY, BILIRUBIN
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50937364
Report Number1823260-2026-04014
Device Sequence Number10722373
Product Code CIG
UDI-Device Identifier07613336235522
UDI-Public07613336235522
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K131544
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number10421436190
Device Lot Number930395
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient SexFemale
Patient RaceAsian
-
-