| Model Number |
BF-1TH190 |
| Medical Device Problem Codes |
No Display/Image (1183); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported no image is showing video, involving an olympus bronchovideoscope.The issue was found during preparation for use and before the patient was in the room.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.In addition, the investigation found a b30 error code due to a faulty plug unit.Based on the results of the investigation, the most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|