Medline Industries, LP MEDLINE; ELECTRODE,ECG,TAB,10/CARD,5000/CS, ADULT Back to Search Results
Catalog Number MDSM616201
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the electrodes do not stick.
 
Additional Manufacturer Narrative
Customer dawn reported that the electrodes do not stick and the defect was documented as adhesive weak/won't stick.At this time, a sample(s) was not received for investigation; however, a lot number/serial number was recorded, and the investigation included review of the lot/serial number and recent complaint trends/scar activity for the past 3 months.The complaint was not confirmed due to the fact that this issue can only be confirmed by further investigating the sample or a provided photo; the cause was documented as sample required to determine cause.No further action was taken because objectionable trends were not seen for this issue, division has been informed and will continue to monitor trends, and there was no reported death or serious injury associated with this event.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ELECTRODE,ECG,TAB,10/CARD,5000/CS, ADULT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rahul kota
three lakes drive
northfield, IL 60093--275
MDR Report Key50937407
Report Number1417592-2026-53755
Device Sequence Number21207012
Product Code DRX
UDI-Device Identifier40888277352330
UDI-Public40888277352330
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K093327
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDSM616201
Device Lot Number85923010003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2023
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-