• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Respironics, Inc. Bipap A40 Pro; Ventilator, Continuous, Minimal Ventilatory Support, Facility Use

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. Bipap A40 Pro; Ventilator, Continuous, Minimal Ventilatory Support, Facility Use Back to Search Results
Model Number FRX3100S14
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a ventilator inoperative error code was found on a bipap 40 pro device.The device was not in use on a patient at the time of the occurrence.During the evaluation of the device at third-party service center, the third-party service technician observed the error code, the power fail voltage is outside its valid range of 4.775 to 5.675 for at least ten seconds, in the device's error log.The third-party service technician further evaluated but was not able to determine the cause of the issue for this device.The root cause is not confirmed at this time.The device was scrapped per the customer request.A final report is being submitted.If new information becomes available following the completion of the device repair that changes the outcome of the investigation and associated medical device reporting a supplemental report will be completed.Additionally, during the service of the device, the system printed circuit assembly needs replaced for another error code, error high internal o2, that was observed on the device's error log.This was unrelated to the reportable malfunction/issue.
 
Additional Manufacturer Narrative
The reported failure of ventilator inoperative is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.A field action was initiated with record id "[fsn-2023-cc-src-039]", and consisted of a field safety notice.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device exceeds expected life so there is no dhr required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bipap A40 Pro
Common Device Name
Ventilator, Continuous, Minimal Ventilatory Support, Facility Use
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50937488
Report Number2518422-2026-046582
Device Sequence Number15886450
Product Code MNT
UDI-Device Identifier00606959055100
UDI-Public(01)00606959055100
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberFRX3100S14
Device Catalogue NumberFRX3100S14
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-