MEDTRONIC PUERTO RICO OPERATIONS CO. InterStim; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON Back to Search Results
Model Number 97810
Medical Device Problem Codes Failure to Interrogate (1332); Malposition of Device (2616); Communication or Transmission Problem (2896); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from a healthcare provider (hcp) and a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stim and gastrointestinal/ pelvic floor.It was reported that they were trying to put the device into mri mode, but they were getting "device not responding".The caller stated that the patient indicated they hadn't used the implant or external equipment for a long time.The agent walked the caller through repositioning the communicator but they continued to receive "device not responding." the agent walked the caller through using the recharger and they described seeing open loop charging screen with the battery in the red and antenna locate numbers.The caller repositioned the recharger multiple times and the hi ghest value they were able to obtain was 15 and noted recharger was warm.While working with the recharger, the agent heard the patient in the background say "it hurts." the agent asked if the caller could palpate the implant and they were not able to feel it.The agent reviewed that the caller could continue to reposition the recharger to try and get higher numbers to see if they could obtain better connection to which the implant would start charging normally.The issue was not resolved through troubleshooting.The patient was redirected to their hcp to further address the issue.The patient indicated the implanting hcp was no longer practicing and didn't provide a current managing hcp.The agent transferred the caller to national answering service (nas) to page a representative.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
InterStim
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50937696
Report Number3004209178-2026-16560
Device Sequence Number15886451
Product Code EZW
UDI-Device Identifier00763000203849
UDI-Public(01)00763000203849(17)220528
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2022
Device Model Number97810
Device Catalogue Number97810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured12/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age83 YR
Patient SexFemale
-
-