| Model Number |
72404130 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
07/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this patient with an artificial urinary sphincter (aus) experienced recurring incontinence.A surgical procedure was performed during which the cuff was explanted and replaced with a smaller size cuff.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this patient with an artificial urinary sphincter (aus) experienced recurring incontinence.The physician did not believe the device caused or contributed to the incontinence.A surgical procedure was performed during which the cuff was explanted and replaced with a smaller size cuff.The most likely/suspected cause of the recurring incontinence was either atrophy or incorrect sizing of the implant during original implantation.Additionally, the physician did not believe the cuff was initially placed in the correct location.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72400024
serial number: n/a
batch/lot number: 1100741130
model/catalog description: balloon fgs 61-70 cm
unique identifier (udi) #: (b)(4)
model number/catalog number: 72404127
serial number: n/a
batch/lot number: 1100714367
model/catalog description: control pump fgs iz
unique identifier (udi) #: (b)(4)
b3: date is unknown, so estimated date was entered.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100741130.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 1100714367.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).B3: date is unknown, so estimated date was entered.
|
| |
|
Search Alerts/Recalls
|