Spectranetics Corporation TightRail Rotating Dilator Sheath; Dilator, Vessel, For Percutaneous Catheterization
|
Back to Search Results |
|
| Model Number |
545-513 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A lead extraction procedure commenced to remove a right ventricular (rv) and right atrial (ra) lead due to cied system/pocket infection.Spectranetics lld ez lead locking devices (lld ezs) were inserted into each lead to provide traction.Beginning with a spectranetics 13f tightrail rotating dilator sheath on the rv lead, the tightrail became stuck on the lead and lld ez, requiring intervention by cutting the rv lead and lld ez (mdr #3007284006-2026-00459).Therefore, the rv lead and lld ez remnant remained in the patient (3007284006-2026-00463).The extraction procedure continued, a new tightrail was used, and the ra lead was successfully removed.The patient survived the procedure.This report captures the 13f tightrail that entrapped the lead and lld, requiring intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
A4) patient weight - not available from facility.B6/b7) patient information regarding relevant tests/laboratory data or medical history - not available from facility.H3/h6) the tightrail was not returned, thus no product investigation was performed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Search Alerts/Recalls
|
|
|