| Model Number |
NVRO-25 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Heart Block (4444)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Clinical information: crd (b)(4) - envision ide study, patient site (b)(6).It was reported that on (b)(6) 2026, a 25mm navitor vision valve was chosen for implant using a small flexnav delivery system.A pre-implant balloon valvuloplasty was done with a 20mm non-abbott balloon.The valve got fully re-sheathed 1 time due to the implant being too high.The final implant depth at non-coronary cusp (ncc) was 3.9mm and left coronary cusp (lcc) was 4.8mm.The patient developed a left bundle branch block (lbbb) during the procedure.No treatment was required.The patient was then discharged from the hospital the following day.Subsequent information received:
it was reported that on (b)(6) 2026, the patient returned for the 30 day follow-up and a new onset atrial fibrillation was noted.For treatment, the patient started eliquis.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of heart block was reported intra-procedure.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the available information the cause for the reported heart block cannot be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|