• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Tandem Diabetes Care t:slim X2 Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Tandem Diabetes Care t:slim X2 Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Catalog Number 1014071
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
t:slim X2 Insulin Pump with Control-IQ+ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50937915
Report Number3013756811-2026-177273
Device Sequence Number18575542
Product Code QFG
UDI-Device Identifier00389152407111
UDI-Public(01)00389152407111(22)1017688(11)260331
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K250792
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number1014071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/31/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age37 YR
Patient SexMale
-
-