| Model Number |
P15P-001 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thromboembolism (2654)
|
| Date of Event |
08/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2026 the site contacted to berlin heart clinical affairs to report that a patient, who was supported with an excor pediatric vad system, suffered several strokes with subsequent hemorrhagic conversion.On (b)(6) 2026 an arterial non-cns thromboembolism was identified in spleen, kidney, liver and mesenteric ileus.Small fibrin deposits were noted in the pump during other neurological events.The support was withdrawn and transitioned to palliative care on (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
The blood pump pu valves; 15 ml in/out; ø 9 mm, s/n 2530283 was placed on the patient on (b)(6) 2026 until the pump was explanted on (b)(6) 2026.The event occurred on (b)(6) 2026, which is 78 days after the pump was placed on the patient, who was being supported with an
excor active driving unit.We have reviewed the production records of the excor blood pump.This pump was produced according to our specifications.
|
| |
|