• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: St. Jude Medical - Neuromodulation (Puerto Rico, LLC) Prodigy MRI IPG, 16CH, Rechargeable; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

St. Jude Medical - Neuromodulation (Puerto Rico, LLC) Prodigy MRI IPG, 16CH, Rechargeable; SCS IPG Back to Search Results
Model Number 3772
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced increase in recharge burden.As such, surgical intervention took place wherein the ipg was explanted and replaced to address the issue.Therapy was confirmed post op.
 
Additional Manufacturer Narrative
Date of event (b3) and patient weight (a4) are estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Prodigy MRI IPG, 16CH, Rechargeable
Common Device Name
SCS IPG
Manufacturer (Section D)
St. Jude Medical - Neuromodulation (Puerto Rico, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
St. Jude Medical - Neuromodulation (Puerto Rico, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
+1(972)309-8000
MDR Report Key50938705
Report Number3006705815-2026-06161
Device Sequence Number15886329
Product Code LGW
UDI-Device Identifier05415067017093
UDI-Public(01)05415067017093(10)A000060272(17)200618
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2020
Device Model Number3772
Device Catalogue Number3772
Device Lot NumberA000060272
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/19/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS Lead
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight140 LB
-
-