• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU Medical Healthcare Manufacturing S.A. de C.V. Bivona Flextend Pediatric and Neonatal Trach Tube; Tube, Tracheostomy (W/Wo Connector)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU Medical Healthcare Manufacturing S.A. de C.V. Bivona Flextend Pediatric and Neonatal Trach Tube; Tube, Tracheostomy (W/Wo Connector) Back to Search Results
Catalog Number 67PFSS35
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bivona Flextend Pediatric and Neonatal Trach Tube
Common Device Name
Tube, Tracheostomy (W/Wo Connector)
Manufacturer (Section D)
ICU Medical Healthcare Manufacturing S.A. de C.V.
ave calidad no. 4, parque
industrial int'l tijuana
tijuana NA 22425
MX  22425
Manufacturer Contact
reed covert
951 calle amanecer
san clemente, IL 92673
+1(224)706-2300
MDR Report Key50939873
Report Number9617604-2026-01110
Device Sequence Number13451850
Product Code JOH
UDI-Device Identifier10351688518682
UDI-Public(01)10351688518682
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number67PFSS35
Device Lot NumberUnknown
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight19 LB
-
-