The customer reported leakage on the item and stated that the stopcock was leaking out the bottom, with an issue with the turning valve itself.The packs production record was reviewed and there were no line down situations, deviations, run-with outs, or non-conformance reports that might have had an impact on this issue; two samples were returned for investigation, and the two returned samples leaked upon aspiration.Based on the investigation results, the reported defect is confirmed.Root cause can be attributed to vendor design deviation, and the overall conclusion was that the issue was related to an error in the component manufacturing process.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|