As reported, the patient had an abdominal aortic pseudoaneurysm (diameter: 50 mm).During the procedure, a 5f angiographic catheter was used to deliver the first azur 20-30 coil.Significant resistance was encountered during advancement, resulting in separation of the delivery wire from the coil.The angiographic catheter was subsequently withdrawn.A new angiographic catheter was then used, and two azur 20-30 coils were successfully delivered without issue.A covered stent was subsequently deployed, and the procedure was completed successfully.The patient returned safely to the ward, and no harm or adverse clinical impact occurred.Patient condition, stable.
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device is reported available for return to the manufacturer for analysis.At this time the alleged product issue cannot be confirmed.If the device or additional information is received, microvention, inc., will submit a supplemental report.The instructions for use (ifu) identifies premature coil detachment as potential complications associated with use of the device.
|