The customer reported that during a patient call, the autopulse nxt platform (sn (b)(6) stopped delivering compressions.The customer stated that it is unknown whether the nxt platform displayed any alert when it stopped delivering compressions.The customer mentioned that the nxt platform ventilation openings were not blocked by clothing or the patient, but there was a hygiene barrier on the platform.The customer suspected that the platform's hygiene barrier may not have allowed sufficient airflow to the fan and that the batteries may have become hot.The customer didn't report any complaint related to the batteries used during the reported event and stated that the batteries have been functioning properly since then.The customer did not report any harm, injury, or death and stated that there was minimal impact on the patient, as they just resumed manual cpr.
|