Covidien LigaSure XP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number LXMJ23S
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during lobectomy procedure, the device was connected to the generator and output sound was heard; however, it could not be confirmed that output was being delivered in the surgical field.The device had no seal with no evidence of energy delivery and end tones heard. a new product was connected to the same generator and the first sealing was performed without any issues on approximately 8 to 10 millimeter in the interlobar / interlobar region. there was no patient injury.
 
Event or Problem Description
Since the output could not be confirmed, the handle was opened without waiting for the seal completion sound (because it was released from the switch), which caused a re-grasp alert.
 
Additional Manufacturer Narrative
D10 concomitant product: vlft10gen, vlft10gen ft series energy platformx1 (serial no.# unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LigaSure XP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
Covidien
6044 longbow court
boulder CO 80301 -329
Manufacturer (Section G)
Covidien
6044 longbow court
boulder CO 80301 -329
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112--439
+1(800)646-4633
MDR Report Key50941719
Report Number1717344-2026-01334
Device Sequence Number10722587
Product Code GEI
UDI-Device Identifier10884521830929
UDI-Public(01)10884521830929(17)310311(10)60710112X(20)02
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K223158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLXMJ23S
Device Catalogue NumberLXMJ23S
Device Lot Number60710112X
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/23/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/30/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11