| Model Number |
RD900AEU |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A healthcare facility in seattle reported that the rd900aeu neopuff infant resuscitator failed the valve system test during device inspection, the pressure was observed to reach 83 cmh2o, exceeding the maximum pressure range indicated on the manometer.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
|
| |
|