The customer reported that 1 out of the 6 glucose meters ordered were defective and incorrectly reading glucose levels, and when level 2 test strips are entered, the reading should be between 74-99 and this item kept having a reading of 71.Medline received 1 evencare proview blood glucose meter in like-new condition with working batteries; test strips and solution used by the account were not sent for investigation, and the sample was tested according to the instructions for use using control evencare proview test strips and evencare proview control solutions.After testing was conducted, the meter displayed glucose readings within both control solution ranges, 71 mg/dl for level 2 and 225 mg/dl for level 3, and medline experienced no issues during testing; the meter turned on and off and displayed readings within control solution range.The complaint was not confirmed with the received sample, and based on review of the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications; the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|