• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Dexcom, Inc. Dexcom G7 15 Day Continuous Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Dexcom, Inc. Dexcom G7 15 Day Continuous Glucose Monitoring System Back to Search Results
Model Number 9510-007
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/11/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred frequently between 9/10/2026 and 9/11/2026.The sensor was inserted into the arm on (b)(6) 2026.It was indicated that the patient entered bg values outside the 40¿400 mg/dl (2.2¿22.2 mmol/l) range, which is misuse of the device.Data was provided for investigation.The allegation was confirmed due to the finding of inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred frequently within the investigation window.The probable cause of the inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter could not be determined.The reported glucose values fall within the c zone of the parkes error grid.The data investigation found a difference in values that fall within the d zone of the parkes error grid.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Dexcom G7 15 Day Continuous Glucose Monitoring System
Common Device Name
Continuous Glucose Monitor
Manufacturer (Section D)
Dexcom, Inc.
6340 sequence dr.
san diego CA 92121
Manufacturer (Section G)
Dexcom, Inc.
232 s dobson rd
mesa AZ 85202
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
+1(858)200-0200
MDR Report Key50941913
Report Number3004753838-2026-263382
Device Sequence Number18828707
Product Code QBJ
UDI-Device Identifier00386270004758
UDI-Public(01)00386270004758(11)260501(17)271031(10)1726146006
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9510-007
Device Catalogue NumberSTP-FT-011
Device Lot Number1726146006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/29/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/01/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Omnipod 5
Patient Age33 YR
Patient SexUnknown
Patient Weight195 LB
-
-