Biosense Webster Inc QDOT MICRO; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation Back to Search Results
Catalog Number D139505
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 09/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device number lot 31928212l and no internal actions related to the complaint were found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition.If the product is received at a later date, the investigation will be updated as applicable.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a premature ventricular contraction ablation procedure with a qdot micro catheter and the patient experienced a cardiac tamponade which required pericardiocentesis.After the ablation, during hemostasis, st elevation occurred, and echocardiography revealed pericardial effusion.The patient recovered after drainage of the pericardial fluid.The patient has fully recovered.No error message observed on the biosense webster equipment during the procedure.Transseptal puncture was not performed.No ablations performed within the pulmonary veins.No evidence of a steam pop.The flow setting was set to pre:2sec, post: 0sec.The correct catheter settings were selected on the generator.No issues with flow rate change at the start of the ablation.Physician¿s assessment of health hazard is non-serious (moderate/minor).There were no abnormalities observed during product use.The physician¿s opinion on the cause of this adverse event is that upon post-procedural review of the fluoroscopy images, the physician observed that cardiac motion appeared to worsen around the time of coronary sinus (cs) catheter insertion.Therefore, the physician believes that the event was more likely related to inadvertent contact or abrasion of the myocardial tissue during cs catheter insertion, rather than the event being attributable to the ablation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO
Common Device Name
Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation
Manufacturer (Section D)
Biosense Webster Inc
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
Biosense Webster Inc (Juarez)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
katie seppa
31 technology dr
irvine, CA 92618
+1(513)337-8052
MDR Report Key50941987
Report Number3013300026-2026-03214
Device Sequence Number10735158
Product Code OAE
UDI-Device Identifier10846835016758
UDI-Public(01)10846835016758(17)290419(10)31928212L
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P210027
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberD139505
Device Lot Number31928212L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/20/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient SexFemale
Patient RaceAsian
-
-