The actual date of event is unknown.A device history record, complaint history review, and risk review on failure modes are performed on each device with reportable malfunction(s) along with other methods of investigation as coded in section h6 of this mdr report.Investigation summary:
the reported "display issue" event was confirmed and attributed to a defective logic board.An external and internal inspection of the suspect pcu was performed, and it was observed that the logic board was found to be functionally defective during internal inspection; however, no signs of physical damage were observed.A temporary replacement of the suspect pcu logic board was performed for testing purposes only.The returned suspect pcu passed all the pm (preventive maintenance) tests in alaris system maintenance (asm) (v12.5).The facility reported a display issue; no event date or time was provided.A review of the logs was performed, and no records related to the reported event were found.The last infusion recorded on the suspect pcu occurred on june 22, 2026, at 8:02 pm, during which a pump module with s/n 18187073 was programmed with a rate of 125 ml/h and a vtbi of 256.58 ml.Bd alaris¿ system with guardrails¿ suite mx (12.3) states:
do not use a device that appears to be damaged.Send the device to biomedical engineering for repair.The device was in use for treatment purposes as intended per 21 cfr 820.198(d)(2).Root cause:
the root cause of the reported 'display issue' event was determined and attributed to a defective logic board.Per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
|