| Model Number |
9510-007 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that signal loss occurred.Performance data was reviewed.The allegation was confirmed due to the finding of signal loss over an hour within the investigation window.The probable cause was the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Search Alerts/Recalls
|