| Model Number |
CF-HQ190L |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed
a root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure
olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities
date of event is unknown.Additional information will be provided once available.
|
| |
|
Event or Problem Description
|
|
The customer reported the colonovideoscope had communication error.During the initial evaluation of the device, no image and error code b30 (scope communication error) were observed.No patient harm was reported.
|
| |
|