| Model Number |
381121-19 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that universal surgical manipulator (usm) 2 faulted with error 32056 while the site's clinical sales representative (csr) was troubleshooting video issues.The intuitive surgical inc technical support engineer (tse) reviewed the system logs and confirmed the reported error.The tse had the csr perform a hard reboot, with no change.The caller stated that the site has several cases scheduled for the following day that they would need all 4 usms for.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the universal surgical manipulator (usm) 2.The system was verified and ready for use.Intuitive surgical, inc.(isi) has received the unit; however, failure analysis is still ongoing.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
|
| |
|