• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC daVinci Xi; PATIENT SIDE CART, 4-ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC daVinci Xi; PATIENT SIDE CART, 4-ARM Back to Search Results
Model Number 380652-53
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to the start of a da vinci-assisted procedure, following anesthesia and prior to port placement, the customer experienced recurrent recoverable errors 1105 and 1162 on universal surgical manipulator (usm) arm#4.Upon reviewing the logs, the intuitive surgical inc.(isi) technical support engineer (tse) determined that usm#4 was faulty and required replacement.The tse guided the caller to disable arm#4 and advised proceeding with the three remaining arms, which the customer indicated would most likely be sufficient for the upcoming procedure.However, the case was ultimately converted to open surgery.No patient injury was reported as a result of the issue or the conversion.
 
Additional Manufacturer Narrative
An intuitive field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Based on the field evaluation, this reported event was confirmed.Fse replaced the universal surgical manipulator(usm)#4.The system was tested and verified as ready for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
daVinci Xi
Common Device Name
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50942417
Report Number2955842-2026-40386
Device Sequence Number10740005
Product Code NAY
UDI-Device Identifier00886874110720
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Nurse
Remedial Action Other,Notification
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number380652-53
Device Catalogue Number380652
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
ISIFA2023-07-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Patient SexUnknown
-
-