| Model Number |
381121-40 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to start of da vinci-assisted benign hysterectomy surgical procedure, site contacted intuitive technical support engineer (tse) to report integrated electrosurgical unit (iesu) powered off during preparation of procedure.The customer turned on iesu.There was activation sound continuously.Tse reviewed event logs via onsite.M 19, m 36 were recorded.The procedure was continued with force triad.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the integrated electrosurgical unit (iesu).The system was verified and ready for use.Intuitive surgical, inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.
|
| |
|