| Model Number |
380652-56 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to the start of a da vinci-assisted surgical procedure, the user informed the technical support engineer (tse) that a repeated recoverable fault 32056 occurred on the universal surgical manipulator (usm3).The user disabled usm3 and continued the procedure using the remaining three usms without further issues.No known impact or patient consequence was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Fse replaced universal surgical manipulator (usm3) due to error 32056.The system was tested and verified as ready for use.The complaint was confirmed based on field evaluation.Isi has not received the usm3 for evaluation.Therefore, no product evaluation was performed.The root cause of error 32056 points to axes controller, arm (aca) in usm3 failed to initialize i2c device.
|
| |
|