| Model Number |
471049-11 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after a da vinci-assisted surgical procedure, a cadiere forceps instrument was observed not being manipulated.The procedure was completed with no reported injury.Intuitive surgical followed up with the initial reporter and obtained the following additional information: no fragment falling into the patient.No fragments retrieved from the patient, nor patient returned to hospital experiencing any post-surgical complications.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The cadiere forceps instrument was analyzed and found to have a broken main tube.The cadiere forceps was found to have a broken main tube approximately zero inches from the distal tip.There is missing material from the broken main tube.Components adjacent to this broken main tube show damage which may indicate potential mishandling/misuse.The proximal clevis is dislodged as a result.The probable root cause is attributed to damage during use, which may result from an accidental drop of the instrument or inadvertent collisions with other instruments.Excessive side loading on the instrument can also cause the main tube wall to collapse inwards leading to cracking and subsequent breakage.This issue can be resolved by using an alternate instrument to complete the procedure.
|
| |
|