| Model Number |
380652-51 |
| Medical Device Problem Codes |
No Device Output (1435); Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the universal surgical manipulator (usm) 1 had a non-recoverable fault 32056 when the system powered on.The customer power cycled the system five times, but the issue remained.The intuitive surgical, inc.(isi) technical service engineer had the customer hard power cycle the system, but the error returned again.There was no report of patient involvement.Isi contacted the clinical territory associate and obtained the following information: the issue was identified when setting up the operating room (or) in the evening on the 6th to get the system ready for the next day.The two patients (procedures) were cancelled and are being rescheduled.No procedure was going on at the time of the event.The robot was isolated until a field service engineer (fse) was able to go on site.There was no delay involved, just cancellation of upcoming procedures.There was no injury to any patient.The fse resolved the issue on the afternoon of the 7th.Isi contacted the technical support engineer and obtained the following information: with error 32056, the customer could have used the system and proceeded with the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The fse was able to reproduce the issue and replaced the universal surgical manipulator (usm) 1.The system was tested and verified as ready for use.As of the date of this report, the usm has not yet been received by intuitive surgical, inc.(isi) for evaluation.
|
| |
|
Search Alerts/Recalls
|