Model Number 470183-19
Medical Device Problem Code Arcing (2583)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive the permanent cautery hook instrument to perform failure analysis (fa).Fa was able to confirm and reproduce the customer reported complaint.The instrument was found to have a dislodged cautery hook at the distal end.The hook is not fully intact at the molded insulator and as a result energy delivery is observed in an unintended location.Smoke was observed coming from the distal clevis during the energy delivery test.The root cause for the failure is not established.
 
Event or Problem Description
It was reported the permanent cautery hook instrument was twisted and loose.There was no report of any patient injury.Intuitive surgical, inc.(isi) followed up with the initial reporter to obtain additional information; however, no further details are available regarding the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EndoWrist
Common Device Name
PERMANENT CAUTERY HOOK
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50942439
Report Number2955842-2026-40384
Device Sequence Number9498722
Product Code NAY
UDI-Device Identifier00886874112311
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470183-19
Device Catalogue Number470183
Device Lot NumberK10260416 0181
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/01/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Patient SexUnknown
-
-