| Model Number |
470183-19 |
| Medical Device Problem Code |
Arcing (2583)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the permanent cautery hook instrument to perform failure analysis (fa).Fa was able to confirm and reproduce the customer reported complaint.The instrument was found to have a dislodged cautery hook at the distal end.The hook is not fully intact at the molded insulator and as a result energy delivery is observed in an unintended location.Smoke was observed coming from the distal clevis during the energy delivery test.The root cause for the failure is not established.
|
| |
|
Event or Problem Description
|
|
It was reported the permanent cautery hook instrument was twisted and loose.There was no report of any patient injury.Intuitive surgical, inc.(isi) followed up with the initial reporter to obtain additional information; however, no further details are available regarding the reported event.
|
| |
|