• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 303, INC. BD Pyxis¿ Anesthesia Station ES; Automated Dispensing Cabinet

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. BD Pyxis¿ Anesthesia Station ES; Automated Dispensing Cabinet Back to Search Results
Model Number 500400001500
Medical Device Problem Codes Application Network Problem (2879); Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ anesthesia station es, multiple machines entered recovery mode and could not be fully rebooted.The customer stated that the machines were disconnected from both the ethernet network and power outlets during downtime in the or space.After reconnecting the ethernet and power connections to or1 and or6, the machines displayed a recovery mode error and failed to complete the startup process.Multiple reset attempts were performed by the technicians; however, the machines remained in recovery mode, preventing further troubleshooting and normal operation.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 20-may-2020 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that station became stuck in windows recovery mode after being powered on and unable to connect to the hospital network because the network cable was connected to the incorrect wall port.The field service engineer (fse) accessed the windows recovery environment and selected the option to bypass signature verification, allowing the pending windows update to complete successfully and reconnected the network cable to the correct network port.The system functioned as intended after the field service engineer (fse) resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ Anesthesia Station ES
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50942481
Report Number2016493-2026-67397
Device Sequence Number15850346
Product Code BRY
UDI-Device Identifier10885403533242
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500400001500
Device Catalogue Number327
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-