| Model Number |
500400001500 |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ anesthesia station es mini matrix pocket 1.6 extended completely outward and exposed all pockets when pulled, regardless of the number of medications in the drawer.The user confirmed when refilling, the drawer extended beyond the last pocket and produced a quiet buzzing noise, indicating that the drawer required further extension.The customer stated that this malfunction occurred while dispensing the medication.There were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 12-feb-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device's mini engine was faulty.A field service engineer (fse) replaced the mini engine to resolve the issue.The system functioned as intended after the field service engineer repaired the device.
|
| |
|