| Model Number |
0998-00-0800-53 |
| Medical Device Problem Codes |
Gas/Air Leak (2946); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) malfunctioned.No harm reported.Unit stopped working while alarming and was unable to troubleshoot.Patient was put on another known good unit to avoid therapy delay.Onsite, fse could not confirm what alarm was alarming during use.Initial power up and run of unit had no issues.However, there was a gas loss alarm along with technical error 118 in the background.As a pre caution the power management board was replaced for error code 118.Unit was then ran for a couple hours and no issues occurred or alarms.Verified unit passed manufacturer specs, unit returned to customer.No other patient information provided.Fse completed repair successfully.Product passed all functional and safety tests per manufacbturer specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter:
(b)(6).Event site name:
(b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|