| Model Number |
160021 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
09/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer states: on (b)(6).2026, during ventilation, the device displayed the error message ¿o¿ sensor failed¿ and stopped displaying the o¿ value.The fio¿ was set to 30% when the message appeared.The patient¿s oxygen saturation subsequently decreased and stabilized at 91%.The fio¿ was then increased to 100%, and the typical sound of oxygen flowing into the device during flushing was immediately heard.The device was replaced.No permanent patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Hamilton medical ag reference number:(b)(4)
investigation ongoing.
|
| |
|