| Model Number |
SC-2318-50 |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
06/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block b3: exact date unknown, event occurred eight weeks prior to the date the manufacturer became aware of the event.Block d.2b; additional applicable product code: qrb.Additional suspect medical device component involved in the event:
model number/catalog number: sc-4318.Batch/lot number: 37686569.Model/catalog description: clik x anchor sterile kit.Unique identifier (udi)#: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was experiencing loss of efficiency from the spinal cord stimulator (scs).It was also stated that the patient's right spinal cord stimulator (scs) lead had migrated two vertebral bodies which was confirmed through xray.The patient had a device optimization was done without success.The patient underwent a lead replacement procedure and the patient was doing well postoperatively.The explanted lead was discarded by the medical facility and will not be returned.
|
| |
|