| Model Number |
500004502000 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medflex 1000, system user was unable to issue medication.Customer reported receiving a "wrong barcode scanned" error when attempting to issue medication.The customer reported that a malfunction took place when the user tried to dispense medication to the patient.However, there were no delay or adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 27-jan-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, a technical support specialist (tss) reviewed the patient's medication order list and confirmed that the medication was present.The tss verified that a transaction had been completed for the medication and discussed the findings with the customer.The tss advised the customer to contact support immediately if the wrong barcode scanned error occurred again so that further investigation could be performed.As the medication was not being actively issued at the time and no ongoing issue was observed, no additional troubleshooting was required and confirmed that there were no issues with the device, and no further investigation was required.The system functioned as intended after the technical support specialist investigated the issue.
|
| |
|
Search Alerts/Recalls
|