Model Number 500001002502
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medstation¿ es auxiliary, the drawer 2 was failed and displayed a device not detected on bus error.Reboot and recovery attempts were performed, but the issue persisted.The customer stated that this malfunction caused a delay in dispensing medication to patients.However, there were no adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 20-jul-2016 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the drawer 2 was not detected on the bus due to a damaged flat flex cable and a loss of communication between the module controller and row boards, resulting in a device not on bus hardware failure.A field service engineer (fse) replaced the damaged flat flex cable, but the communication issues persisted.Fse then replaced the pyxis bus cable.Diagnostic testing confirmed proper communication and functionality, and the customer successfully recovered and tested the drawer and cubies through the application, verified that the issue was resolved.The system functioned as intended after the field service engineer repaired the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedStation¿ ES Auxiliary
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50942731
Report Number2016493-2026-67434
Device Sequence Number9534150
Product Code BRY
UDI-Device Identifier10885403533235
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500001002502
Device Catalogue Number324
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-