| Model Number |
0998-00-0800-53 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, g6,h2,h11
corrected fields : g3
this mdr is being updated based on findings identified during retrospective review under capa.The updated parent investigation for the reported gas loss issue incorporated capa findings and determined that the reported alarm condition could not be reproduced.
|
| |
|
Event or Problem Description
|
|
It was reported that during use the cardiosave intra aortic balloon pump (iabp) displayed gas loss in iab circuit.There was no reported harm or injury.Fse evaluated the unit and noticed that, the compressor was sluggish and slow to come up to pressure.The compressor and the expired safety disk and tidal volume disk were replaced.Unit passed all the tests.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to charactreisation limit e1(initial reporter): everell sinclair
a getinge field service engineer (fse) was dispatched to evaluate the unit, they noticed the compressor is sluggish and is slow to come up to pressure, compressor caused false leak alarms.The fse replaced the compressor and the expired safety disk and tidal volume disk.The unit returned to normal function as well as passed all safety and functional tests.The failure analysis and testing department received the following part associated with this complaint:
compressor scroll
this part was received with a reported unit failure message of gas loss alarm.The failure analysis and testing dept.Performed a visual inspection, and part looks in good condition.Installed compressor scroll into the cardiosave test fixture and tested to the cardiosave service manual.Performed functional tests and passed.Also, pumped the cardiosave test fixture for 3 hours and the fat dept.Did not observe gas loss alarm message on screen.The fat dept.Could not verify the failure message of gas loss alarm.Compressor scroll passed testing.Retaining compressor scroll in the fat dept.Per procedure.
|
| |
|